The value of ultrasound elastography in differential diagnosis of superficial lymph nodes
Yanrong ZHANG phD 1, Qing LV MD 1, Mingxing XIE MD 1, Feixiang XIANG phD 1, Chengfa LU BM 1, Tianwei YAN BM 1, Wei LI BM 1, Hui XU BM 1, Yan HUANG BM 1, Yehua YIN BM 2,
1.Department of Medical Ultrasound, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan 430022, China; 2.Department of Medical Ultrasound, The First Hospital, Jianli 434000, China;
 Download: PDF(344 KB)  
 Export: BibTeX | EndNote | Reference Manager | ProCite | RefWorks
Abstract The study evaluated the value of ultrasound elastography in differentiating the benign and malignant superficial lymph nodes. A total of 112 subjects, including 82 patients with enlarged lymph nodes and 30 healthy volunteers, were recruited. All the subjects were examined by B-mode ultrasonography, power Doppler ultrasonography and elastography. Most of the patients were histopathologically confirmed by needle aspiration cytology and some patients were diagnosed by clinical data combined with follow-up findings. The sensitivity, specificity and accuracy of B-mode ultrasonography were 59.8%, 76.5% and 67.1%, those of the blood flow classification by power Doppler ultrasonography 77.0%, 82.3% and 79.4% and those of elastographic classification 74.7%, 97.1% and 84.5%, respectively. The elasticity of the lymph nodes was quantitatively measured and defined as stiffness value. When the stiffness value of 2.395 was taken as the cutoff point, the sensitivity and specificity of elastography were 78.41% and 98.51%, and the Youden index reached the highest, with the value being 0.7692. The stiffness values of two indeterminate benign lymph nodes, the elastrographic findings of which were rated as patterns 2 and 3, were below the cutoff point. The elastographic findings of 10 malignant lymph nodes were also classified as pattern 2 or 3. Only one of them had the stiffness values below the cutoff point. It was concluded that ultrasound elastography is a novel, noninvasive and convenient tool for the differentiation of the nature of the superficial lymph nodes in clinical practice.
Keywords ultrasound elastography      lymph node      power Doppler sonography     
Issue Date: 05 September 2009
 Cite this article:   
Chengfa LU BM,Yanrong ZHANG phD,Mingxing XIE MD, et al. The value of ultrasound elastography in differential diagnosis of superficial lymph nodes[J]. Front. Med., 2009, 3(3): 368-374.
 URL:  
https://academic.hep.com.cn/fmd/EN/10.1007/s11684-009-0063-0
https://academic.hep.com.cn/fmd/EN/Y2009/V3/I3/368
Ferrari F S, Cozza S, Guazzi G, Della Sala L, Leoncini L, Lazzi S, Stefani P. Roleof Doppler color in the differential diagnosis of benign and malignantadenopathies. Radiol Med (Torino), 1997, 93(3): 242―245
Yang W T, Chang J, Metreweli C. Patients with breast cancer: differences in color Dopplerflow and gray-scale US features of benign and malignant axillary lymphnodes. Radiology, 2000, 215(2): 568―573
Ho S S, Metrewefi C, Ahuja A T. Does anybody know how we should measure Doppler parametersin lymph nodes? Clin Radiol, 2001, 56(2): 124―126

doi: 10.1053/crad.2000.0588
Ophir J, Céspedes I, Ponnekanti H, Yazdi Y, Li X. Elastography:a quantitative method for imaging the elasticity of biological tissues. Ultrason Imaging, 1991, 13(2): 111―134

doi: 10.1016/0161-7346(91)90079-W
Sohn Y M, Kim M J, Kim E K, Kwak J Y, Moon H J, Kim S J. Sonographic elastography combined with conventional sonography: howmuch is it helpful for diagnostic performance? J Ultrasound Med, 2009, 28(4): 413―420
Thitaikumar A, Mobbs L M, Kraemer-Chant C M, Garra B S, Ophir J. Breast tumor classificationusing axial shear strain elastography: a feasibility study. Phys Med Biol, 2008, 53(17): 4809―4823

doi: 10.1088/0031-9155/53/17/022
Saleem S. Elastographyand its role in breast cancer screening. J Coll Physicians Surg Pak, 2009, 19(3): 203―204
Thomas A, Kümmel S, Fritzsche F, Warm M, Ebert B, Hamm B, Fischer T. Real-time sonoelastographyperformed in addition to B-mode ultrasound and mammography: improveddifferentiation of breast lesions? AcadRadiol, 2006, 13(12): 1496―1504

doi: 10.1016/j.acra.2006.08.012
Tardivon A, El Khoury C, Thibault F, Wyler A, Barreau B, Neuenschwander S. Elastographyof the breast: a prospective study of 122 lesions. J Radiol, 2007, 88(5 Pt 1): 657―662

doi: 10.1016/S0221-0363(07)89872-6
Itoh A, Ueno E, Tohno E, Kamma H, Takahashi H, Shiina T, Yamakawa M, Matsumura T. Breast disease: clinicalapplication of US elastography for diagnosis. Radiology, 2006, 239(2): 341―350

doi: 10.1148/radiol.2391041676
Ou B, Luo B M, Feng X, Yang H Y, Zhi H, Wen Y L. The value of ultrasonic elastography in differential diagnosingbreast small masses. Zhonghua ChaoshengYingxiangxue Zazhi, 2007, 16(6): 506―508 (in Chinese)
Zhi H, Xiao X Y, Yang H Y, Wen Y L, Ou B, Luo B M, Liang B L. Semi-quantitating stiffnessof breast solid lesions in ultrasonic elastography. Acad Radiol, 2008, 15(11): 1347―1353

doi: 10.1016/j.acra.2008.08.003
Adler D D, Carson P L, Rubin J M, Quinn-Reid D. Dopplerultrasound color flow imaging in the study of breast cancer: preliminaryfindings. Ultrasound Med Biol, 1990, 16(6): 553―559

doi: 10.1016/0301-5629(90)90020-D
Zhi H, Ou B, Luo B M, Feng X, Wen Y L, Yang H Y. Comparison of ultrasound elastography, mammography, and sonographyin the diagnosis of solid breast lesions. J Ultrasound Med, 2007, 26(6): 807―815
Luo B M, Ou B, Zhi H. Factors affecting the performance of ultrasonic elastographyin breast tumors and its solutions. ZhongguoYixue Kexueyuan Xuebao, 2008, 30(1): 112―115 (in Chinese)
Luo B M, Zeng J, Ou B, Zhi H. Effects ofpressure and compression frequency on diagnosis of breast masses byultrasonic elastography. Zhongguo YixueYingxiang Jishu, 2007, 23(8): 1152―1154 (in Chinese)
Luo B M, Zeng J, Ou B, Zhi H. Effects ofregion of interesting on diagnostic results of ultrasonic elastography. Zhongguo Yixue Yingxiang Jishu, 2007, 23(9): 1330―1332 (in Chinese)
Jank S, Robatscher P, Emshoff R, Strobl H, Gojer G, Norer B. The diagnostic value of ultrasonography to detect occultlymph node involvement at different levels in patients with squamouscell carcinoma in the maxillofacial region. Int J Oral Maxillofac Surg, 2003, 32(1): 39―42

doi: 10.1054/ijom.2002.0326
Dangore S B, Degwekar S S, Bhowate R R. Evaluation of the efficacy of colour Doppler ultrasoundin diagnosis of cervical lymphadenopathy. Dentomaxillofac Radiol, 2008, 37(4): 205―212

doi: 10.1259/dmfr/57023901
Chan J M, Shin L K, Jeffrey R B. Ultrasonography of abnormal neck lymph nodes. Ultrasound Q, 2007, 23(1): 47―54

doi: 10.1097/01.ruq.0000263839.84937.45
Li P, Cai S, Jiang Y X. Ultrasonography differential diagnosis and developmentof superficial lymph node diseases. ZhongguoYixue Yingxiang Jishu, 2007, 23(9): 1409―1413 (in Chinese)
Ghirardelli M L, Jemos V, Gobbi P G. Diagnostic approach to lymph node enlargement. Haematologica, 1999, 84(3): 242―247
Miyagawa T, Tsutsumi M, Matsumura T, Kawazoe N, Ishikawa S, Shimokama T, Miyanaga N, Akaza H. Real-time elastography forthe diagnosis of prostate cancer: evaluation of elastographic movingimages. Jpn J Clin Oncol, 2009, 39(6): 394―398

doi: 10.1093/jjco/hyp026
Kamoi K, Okihara K, Ochiai A, Ukimura O, Mizutani Y, Kawauchi A, Miki T. The utilityof transrectal real-time elastography in the diagnosis of prostatecancer. Ultrasound Med Biol, 2008, 34(7): 1025―1032

doi: 10.1016/j.ultrasmedbio.2007.12.002
Alam F, Naito K, Horiguchi J, Fukuda H, Tachikake T, Ito K. Accuracy of sonographic elastography in the differentialdiagnosis of enlarged cervical lymph nodes: comparison with conventionalB-mode sonography. Am J Roentgenol, 2008, 191(2): 604―610

doi: 10.2214/AJR.07.3401
Dighe M, Bae U, Richardson M L, Dubinsky T J, Minoshima S, Kim Y. Differential diagnosis of thyroid nodules with US elastographyusing carotid artery pulsation. Radiology, 2008, 248(2): 662―669

doi: 10.1148/radiol.2482071758
Uchida H, Hirooka Y, Itoh A, Kawashima H, Hara K, Nonogaki K, Kasugai T, Ohno E, Ohmiya N, Niwa Y, Katano Y, Ishigami M, Goto H. Feasibility of tissue elastographyusing transcutaneous ultrasonography for the diagnosis of pancreaticdiseases. Pancreas, 2009, 38(1): 17―22

doi: 10.1097/MPA.0b013e318184db78
Kolokythas O, Gauthier T, Fernandez A T, Xie H, Timm B A, Cuevas C, Dighe M K, Mitsumori L M, Bruce M F, Herzka D A, Goswami G K, Andrews R T, Oas K M, Dubinsky T J, Warren B H. Ultrasound-based elastography:a novel approach to assess radio frequency ablation of liver massesperformed with expandable ablation probes: a feasibility study. J Ultrasound Med, 2008, 27(6): 935―946
Salomon G, Köllerman J, Thederan I, Chun F K, Budäus L, Schlomm T, Isbarn H, Heinzer H, Huland H, Graefen M. Evaluationof prostate cancer detection with ultrasound real-time elastography:a comparison with step section pathological analysis after radicalprostatectomy. Eur Urol, 2008, 54(6): 1354―1362

doi: 10.1016/j.eururo.2008.02.035
Furukawa M K, Kubota A, Hanamura H, Furukawa M. Clinicalapplication of real-time tissue elastography to head and neck cancer—evaluationof cervical lymph node metastasis with real-time tissue elastography. Nippon Jibiinkoka Gakkai Kaiho, 2007, 110(7): 503―505

doi: 10.3950/jibiinkoka.110.503
[1] Xianxian Li, Hui Xing, Lin Li, Yanli Huang, Min Zhou, Qiong Liu, Xiaomin Qin, Min He. Clinical significance of para-aortic lymph node dissection and prognosis in ovarian cancer[J]. Front Med, 2014, 8(1): 96-100.
[2] Tian Wang, Yan Li, Abidan Tuerhanjiang, Wenwen Wang, Zhangying Wu, Ming Yuan, Shixuan Wang. Correlation of Twist upregulation and senescence bypass during the progression and metastasis of cervical cancer[J]. Front Med, 2014, 8(1): 106-112.
[3] Yang YU, Jibin LIU, Lixue YIN. Application progress of lymphography in oncology[J]. Front Med Chin, 2009, 3(1): 13-19.
Viewed
Full text


Abstract

Cited

  Shared   
  Discussed