Studies on injector pump-ultrasonic nebulizer feeding sample system for MPT-AES |
Feng Guodong, Han Songbai, Yu Aimin |
Department of Analytical Chemistry, College of Chemistry, Jilin University, Changchun 130023, China; |
|
|
Abstract The feeding-sample system used by microwave plasma torch atomic emission spectroscopy (MPT-AES) is the pneumatic nebulization system; its efficiency, however, is not good. A newly built injector pump ultrasonic nebulizer combination feeding-sample system has been designed. Its performance was tested and compared with that of the pneumatic nebulization system. It can be concluded that the newly built feeding-sample system can increase the spectral line intensity by about two to three times and decrease the detection limit by about 2 to 10 times. Moreover, this newly built system can reduce the time taken washing the sample cell from 30 min or so to about 10 min.
|
Issue Date: 05 June 2006
|
|
|